Measuring lens centration means determining how far the optical center is offset from the geometric center. A measurement taken at only one position describes one point on the circumference; it cannot represent the complete circle. The lens must therefore rotate so that the optical surface can be measured at multiple angles and the true center deviation can be calculated.
Why must a lens be rotated for measurement?
Measuring center deviation is fundamentally the process of finding the offset between the lens’s optical center and geometric center. One measurement position only reflects one point. To calculate the actual center deviation, the lens must rotate with the stage while surface data are acquired at multiple angular positions around the complete circumference.
There is one critical assumption: the measurement system assumes that the rotary stage rotates stably around a fixed axis.
If the stage shifts slightly while carrying the lens, the acquired signal contains two parts: the centration error of the lens itself plus the motion error introduced by the stage. The final result then represents the complete measurement system, not the lens alone.
What exactly is rotary-stage runout?
The difference between the stage’s actual motion and its ideal rotation is the runout that must be evaluated. Runout can be divided into three components.
Radial runout
Radial runout is position variation perpendicular to the axis of rotation. It mainly interferes with lens center-deviation measurement and introduces an apparent eccentricity that belongs to the stage rather than the lens.
Axial runout
Axial runout is position variation along the rotation axis. It mainly affects tilt-angle measurements and centration measurements related to surface form or curvature, producing a false tilt error.
Tilt error motion
Tilt error motion is angular error relative to the mean axis of rotation. It causes the measured spot trajectory to depart from an ideal concentric circle and introduces false eccentricity or surface-form signals.
Where does rotary-stage runout come from?
A rotary stage’s accuracy is determined by several factors.
1. Component machining
The machining accuracy of the spindle, stage face, mating surfaces, and other mechanical parts directly affects runout.
2. Assembly
Assembly errors such as coaxiality and perpendicularity also contribute to final stage error.
3. Bearings and support structure
Bearing type, manufacturing precision, stiffness, and operating condition all affect error motion.
4. Actual operating condition
Load, speed, temperature, and the surrounding operating environment can change the measured runout.
How is rotary-stage runout measured?
For very small radial and axial motion, a common method uses a non-contact capacitive or inductive displacement probe aligned with a measurement point on the stage. The probe records the displacement change while the stage rotates.
Principle: A precision-calibrated reference sphere is mounted at the center of the rotary stage, and the capacitive or inductive probe is aligned to the sphere. As the stage rotates, the sensor’s voltage change reflects small radial or axial displacement of the spindle.
Application: High-precision inspection of radial runout and axial motion, with resolution reaching the nanometre level.
Characteristics: Extremely high resolution, suitable for ultraprecision measurement, but highly sensitive to temperature, airflow, and the quality of the measured surface.
- Align the non-contact capacitive or inductive probe to the measurement point and acquire displacement change during stage rotation.
- Rotate the stage through one complete revolution while the sensor continuously records displacement at that point. Smooth rotation produces a small reading variation; radial or axial runout causes the reading to rise and fall with rotation.
- Use the maximum and minimum readings over one revolution; their difference is the total indicator reading under that measurement condition.
Total indicator reading (TIR)
TIR = maximum reading − minimum reading
For example, if the maximum reading is +2 μm and the minimum reading is −1 μm:
TIR = 2 μm − (−1 μm) = 3 μm
Under that measurement condition, the total reading variation at the measured point during one complete revolution is 3 μm.
TIR describes the total indicated variation under a specific measurement condition; it is not automatically equal to the rotary stage’s pure rotational error. The result can also be affected by the form of the measured surface, mounting eccentricity, sensor noise, thermal drift, and other measurement-system effects.
How should a centration system’s rotary-stage performance be judged?
A centration system performs precision inspection through rotation. The rotary stage not only supports the lens but also carries the workpiece through multiple angles for data acquisition. Its rotational stability therefore directly affects the measurement result.
What to verify during system selection
- Do not evaluate the system only by its nominal center-deviation accuracy.
- Check the rotary-stage specification, load condition, and test conditions associated with that accuracy.
- Review repeatability and the measurement method together with center-deviation accuracy.
- Pay particular attention when inspecting large optical components, heavy lenses, and complex lens assemblies.
Based on industry experience, for a centration system with an overall accuracy specification of approximately 0.2 μm, the rotary-stage accuracy should generally be controlled within approximately 200 nm to support the overall system accuracy.
This is why the rotary stage—which appears simply to “turn the lens”—is an indispensable core metric in precision centration measurement. Its stability determines whether the measured signal represents the lens or a mixture of the lens and the stage.
For the preceding workflow step, see our guide to establishing a mechanical datum before centration measurement. To evaluate a complete system, explore Clotho centration systems or discuss your measurement requirements.
Frequently asked questions
Why does a centration system need a precision rotary stage?
The measurement determines center position from a full circular trajectory. If the stage introduces radial, axial, or angular motion, that movement is combined with the optical signal and increases measurement uncertainty.
What is the difference between radial runout and axial runout?
Radial runout is displacement perpendicular to the rotation axis and can produce false eccentricity. Axial runout is displacement along the rotation axis and can affect tilt and surface-related measurements.
Is TIR the same as rotary-stage error motion?
No. TIR is the total maximum-to-minimum variation measured in a specific setup. It can include reference-sphere error, mounting error, environmental drift, sensor noise, and other effects in addition to stage motion.
Can a 0.2 μm centration system use a stage with 200 nm runout?
The source article presents 200 nm as a general industry rule of thumb for this accuracy class. The final requirement should be based on a complete uncertainty budget and measured under representative operating conditions.